- Conductivity Range :
◠0-20.00, 20-200.0, 200-2000 µS/cm, 2-20.000, 20-200, 200-2000 mS/cm for K=1.0
◠0-10.00, 10-100.0, 100-1000 µS/cm, 1-10.000, 10-100, 100-1000 mS/cm for K=0.5
◠0-2.00, 2-20.0, 20-200, 200-2000 µS/cm, 2-20, 20-200 mS/cm for K=0.1
(Range value change depending upon cell selection)
- Conductivity Resolution
- 0.001, 0.01, 0.1 µS/cm & 0.001, 0.01, 0.1, 1 mS/cm
- TDS Range : 0 to 10, 100, 1000, 10000 ppm, 100ppt, 1000ppt
(Depending on TDS Factor)
- TDS Resolution
- 0.001, 0.01, 0.1 ppm & 0.001, 0.01, 0.1 ppt (1ppt=1000ppm)
- Temperature Range
- -5.0 to 130.0°C
- Temperature Resolution
- 0.1°C
- Temperature Compensation
- Auto / Manual
- Temperature Coefficient
- Six - 0-10.000 % per °C (Adjustable) - One Per Range
- Temperature Calibration
- In-built software and can be calibrated by users
- Number of Range
- Six
- Calibration Standard
- Six Conductivity & TDS Standard - One Per Range
- Calibration Factor
- Six - One Per Range
- Range Selection
- Auto / Manual
- Suitable Cell Constant
- 0.1, 0.5 & 1.0 K
- Adjustable Cell Constant
- 0.09 to 0.11 K, 0.45 to 0.55 K & 0.90 to 1.1 K
- Conductivity to TDS Conversion
- 0.40000 to 1.00000 (Adjustable Factor)
- Reading
- Conductivity at actual temperature and at 25°C and TDS with scrolling display.
- TDS Calibration & Analysis
- In Separate Mode
- Error Messages
- “Wrong Standard / Cellâ€, “Cell Shortâ€, “Cell Openâ€, “Try Upper Rangeâ€, “Try Lower Rangeâ€, “Temperature Probe Error†& “Low Memoryâ€
- Display
- 16 x 2 Line Alphanumeric Backlit LCD.
- Keypad
- Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS.
- Calibration Data Storage
- Non-Volatile Memory which can store, print & view with calibration date & time, cell constant, calibration range, standard conductivity solution, measured conductivity at 25°C, actual temperature, calibration factor, instrument serial number, instrument ID, model number etc.
- Analysis Data Storage : More than 500 analysis data which can be viewed & Print, company name, analysis date & time, record number, *Username, *Sample Name, *Bath Number/ID, calibration data with date & time, conductivity at 25°C and conductivity at actual temperature, % Tempco, conductivity to TDS conversion factor, TDS, instrument serial number, Instrument Make & Model & *Signature etc. Secondly, analysis data can be print online without saving analysis data in Instrument.
(*: Not possible to enter data by instrument keypad but it will be handwritten)
- Searching of data
- With record number by scrolling.
- Report Formation
- As per Non-GLP Compliance
- Printer Interface
- Centronics Parallel Port / Serial (RS-232 C) Port
- Printer
- 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
- Power Supply
- DC Adaptor
- Dimensions:
â— Net Dimension: 22 x 21 x 11 cm (L x W x H)
â— Gross Dimension: 32 x 29 x 23 cm (L x W x H)
- Weight:
â— Net Weight: 1.5 kg
â— Gross Weight: 3.5Â kg