Conductivity/TDS/ Resistivity/Salinity/°C Analyzer

Auto Ranging Conductivity/TDS/Resistivity/Salinity/°C Analyzer

4.8 / 5 · 37 verified reviews

Model : µConCal50 | Make : ANALAB | HSN Code : 90278990

Supplied with Conductivity Cell K = 1.0, Temperature Probe, Electrode Stand consists of Clamp, Base Plate & Rod, Dust Cover, DC Adaptor, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.

Share:
Conductivity Range :
● 0-20.00, 20-200.0, 200-2000 µS/cm, 2-20.000, 20-200, 200-2000 mS/cm for K=1.0
● 0-10.00, 10-100.0, 100-1000 µS/cm, 1-10.000, 10-100, 100-1000 mS/cm for K=0.5
● 0-2.00, 2-20.0, 20-200, 200-2000 µS/cm, 2-20, 20-200 mS/cm for K=0.1
(Range value change depending upon cell selection)
Conductivity Resolution
00.001, 0.01, 0.1 µS/cm & 0.001, 0.01, 0.1, 1 mS/cm
TDS Range : 0 to 10, 100, 1000, 10000 ppm, 100ppt, 1000ppt
(Depending on TDS Factor)
TDS Resolution
0.001, 0.01, 0.1 ppm & 0.001, 0.01, 0.1 ppt (1ppt=1000ppm)
Resistivity Range
20.00 MΩ to 50.00 KΩ, 5.00 KΩ, 500.00 Ω, 50.00 Ω, 0.50 Ω
Resistivity Resolution
0.001
Salinity
0.01 to 80000.0 ppm NaCl
Temperature Range
-5.0 to 130.0°C
Temperature Resolution
0.1°C
Temperature Compensation
Auto / Manual
Temperature Coefficient
Six - 0-10.000 % per °C (Adjustable) - One Per Range
Temperature Calibration
In-built software and can be calibrated by users
Number of Range
Six
Calibration Standard
Six Conductivity & TDS Standard - One Per Range
Calibration Factor
Six – One Per Range
Range Selection
Auto / Manual
Suitable Cell Constant
0.1, 0.5 & 1.0 K
Adjustable Cell Constant
0.09 to 0.11 K, 0.45 to 0.55 K & 0.90 to 1.1 K
Conductivity to TDS Conversion
0.40000 to 1.00000 (Adjustable Factor)
Reading
Conductivity at actual temperature and at 25°C, TDS, Resistivity & Salinity with scrolling display.
TDS Calibration & Analysis
In Separate Mode
Error Messages
“Wrong Standard / Cell”, “Cell Short”, “Cell Open”, “Try Upper Range”, “Try Lower Range”, “Temperature Probe Error” & “Low Memory”
Display
20 x 4 Line Alphanumeric Backlit LCD.
Keypad
Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS.
Calibration Data Storage
Non-Volatile Memory which can store, print & view with calibration date & time, cell constant, calibration range, standard conductivity solution, measured conductivity at 25°C, actual temperature, calibration factor, instrument serial number, instrument ID, model number etc.
Analysis Data Storage
More than 750 analysis data can be stored with corresponding calibration data with company name, analysis date & time, record number, batch number / ID, sample name, username, conductivity at 25°C and conductivity at actual temperature, % Tempco, conductivity to TDS conversion factor, TDS, resistivity, salinity, instrument serial number, instrument ID, model number etc. Secondly, analysis data can be print online without saving analysis.
Searching of data
With record number, batch number, Date and by scrolling
Report Formation
As per GLP Compliance
Password Protection
Date & Time, Calibration Mode, Analysis Mode & Delete Stored Analysis Data.
Printer Interface
Centronics Parallel Port / Serial (RS-232 C) Port
Printer
40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
Power Supply
DC Adaptor
Dimensions:
● Net Dimension: 22 x 21 x 11 cm (L x W x H)
● Gross Dimension: 32 x 29 x 23 cm (L x W x H)
Weight:
● Net Weight: 1.5 kg
● Gross Weight: 3.5 kg