- Temperature Controller
- Microcontroller Based
- Detection Principal of Melting Point
- Automatic by Silicon Photodiode
- Heating Media
- By Aluminum Block
- Temperature Range
- +5°C above ambient to 400°C
- Temperature Accuracy :
A) +5°C above ambient to 100°C ……… ±0.2°C
B) 101°C to 200°C …......………………………………… ±0.4°C
C) 200°C to 400°C ………......….…………………… ±0.5°C
- Readability
- 0.1°C
- Overheating Protection
- In-built - Set by user in method parameter
- Heating Before EMP* : Automatic heating depending upon temperature difference
(* EMP: Expected Melting Point)
- Heating Rates
- 0.1 to 20.0ºC/min (Increment of 0.1ºC) - Adjustable
- Maximum Heating Time
- Around 6 minutes from 50°C to 350°C
- Maximum Cooling Time
- Around 6 minutes from 350°C to 50°C
- Detection of Melting :
A) Melting Point
B) Melting Range
C) % Level of Melting Point between Melting Range
- Type of Detection
- Auto / Manual (Color sample can be analysed by Manual detection)
- Melting Point Average Reading
- For three same sample
- Melting Capillary Ports
- Three
- Melting Channel
- Three different sample analysis of different temperature in single run
- Detection of Boiling Point
- Manually by viewing in TFT Display
- Boiling Tube
- One
- Calibration
- Eight Calibration Standards (Six as per Pharmacopeia AND / OR Customer's known standards)
- Method
- 40 methods for Melting sample, 10 for Boiling sample and 1 for unknown sample with View, Print & Delete facility.
- Temperature Sensor
- Duplex PT-100, One for internal use and One for calibration purpose by external Source.
- Glass Melting Capillary
- OD: 1.4 to 1.6mm | ID: 1.0mm | Length: 100mm | One End Sealed
- Sample Filling Height
- 3mm
- Magnification
- 5x
- Recording of Melting Capillary
- Camera with TFT Display for viewing and Recording of Melting Capillary in CAPillaryVIEW. Individual recording storage for each Capillary. (Last Analysis)
- Display
- 20 x 4 Line Alphanumeric Backlit LCD
- Keypad
- Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
- Calibration Data Storage
- Non-Volatile Memory which can store, print & view company name, standard name, standard MP, observed MP, Calibration date and time, last calibration date & time, calibration done by, calibration status, instrument ID, serial number, model number etc.
- Analysis Data Storage
- More than 1500 analysis data can be stored with corresponding calibration data, analysis date & time, record number, Batch number / ID, sample name, username, assigned method number, MP1 (On set temperature) & MP2 (Clear temperature) reading of all three reading, instrument ID, serial number, model number, make, etc. Secondly, analysis data can be print online.
- Searching of data
- With record number, batch number/ID, date, date range and by scrolling.
- Report Formation
- As per GLP Compliance
- 21 CFR Part-11 Compliance :
● In-built with system for ease of operation without PC.
● 2 Admin, 10 Reviewers, 40 Users & 1 Manufacturer with individual Login ID & Passcode.
● Audit trail can store more than 5000 lines in instrument.
● Audit trail includes action taken, date & time, action taken by whom.
● Audit trail can be transfer to PC through ANALABsafe software and can print on 40 / 80 column dot matrix printer for record.
● Admins can set up operation permission of each Users & Reviewers.
● Strict Admins control to audit and monitor usage and access.
● Audit trail record can be printed with Date, Date Range and by individual User’s ID.
● Once Audit trail transfer to the PC or Printer, system will delete Audit trail records and before final deletion, the system will give warning twice.
● Data integrity with database capability to store Analysis Data, Calibration Data, Calibration and Analysis Method, Audit trail and Passcode in instrument.
● No direct access to data from outside the instrument, thereby creating a closed system.
● Audit traceability to ensure that the device and results are appropriately handled.
● Auto User’s logout after defined Logout time.
● Auto Users & Reviewers block when password is entered incorrectly 3 consecutive times
- Printer Interface
- Centronics Parallel Port / Serial (RS-232 C) Port
- Printer
- 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
- PC Interface
- USB Type - B
- PC Connectivity
- By USB Interface Cable with Driver, ANALABsafe software to transfer Analysis data in GLP format, Calibration data and Audit trail from Instrument to PC in non-editable format (Cannot Delete)
- Power Supply
- 230 VAC ±10%, 50 Hz
- Dimensions:
● Net Dimension: 41 x 25 x 22 cm (L x W x H)
● Gross Dimension: 48 x 33 x 35 cm (L x W x H)
- Weight:
● Net Weight: 8 kg
● Gross Weight: 10 kg