- pH Range
- -1.00 to 15.00 pH
- Resolution
- 0.01 pH
- Relative Accuracy
- ±0.02 pH
- mV / Accuracy
- ±2000.0 mV / 0.2mV OR ± 0.05% of reading whichever is greater.
- Slope
- 80 - 120 %
- Auto Buffer Recognition
- 7.00, 4.01 / 9.18 pH @25ºC for Two points
- Display Reading
- Simultaneous - pH, mV & Temperature.
- Buffer Standard
- US / NIST / EURO / DIN
- Short Link
- Provided to diagnose the problem either with Instrument OR Electrode.
- Conductivity Range
- 0-20.00, 20-200.0, 200-2000 µS/cm, 2-20.000, 20-200 mS/cm
- Conductivity Resolution
- 0.001, 0.01, 0.1 µS/cm & 0.001, 0.01 mS/cm
- TDS Range
- 0 to 10, 100, 1000, 10000, 100000 ppm
- TDS Resolution
- 0.001, 0.01, 0.1 ppm & 0.001, 0.01 ppt (1ppt=1000ppm)
- Number of Range
- One
- Calibration Standard
- One
- Calibration Factor
- One
- Suitable Cell Constant
- 1.0 K
- Adjustable Cell Constant
- 0.90 to 1.1 K
- Conductivity to TDS Conversion
- 0.40000 to 1.00000 (Adjustable Factor)
- Display Reading
- Conductivity at actual temperature and at 25°C and TDS with scrolling Display.
- Temperature Range
- 0 to 130.0ºC
- Relative Accuracy
- ±0.1ºC
- Temperature Compensation
- Auto / Manual
- Temperature Calibration
- In-built software and can be calibrated by users
- Error Message
- “Bad Buffer / Electrodeâ€, In-complete Calibration“ pH Over Rangeâ€, “Wrong standard / Cellâ€, “Cell Shortâ€, “Cell Openâ€, “Try Upper Rangeâ€, “Try Lower Rangeâ€, “Temperature Probe Error†& “Low Batteryâ€
- Display
- 16 x 2 Line Alphanumeric Backlit LCD.
- Keypad
- Membrane waterproof polyester soft keypad compliance to UL/ROHS.
- Calibration Data Storage
- Non-Volatile Memory can store & view calibration details.
- Special Features
- Auto Power Off, Backlit On / Off for battery conservation.
- Power Supply
- Rechargeable Battery with Charger
- Dimensions:
â— Net Dimension: 21 x 11 x 5 cm (L x W x H)
â— Gross Dimension: 43 x 11 x 34 cm (L x W x H)
- Weight:
â— Net Weight: 0.5 kg
â— Gross Weight: 2.0Â kg