- Element Range
- Na, K, & Li : 1-200 ppm, Ca : 15-300 ppm, Ba : 30-3000 ppm
- Sensitivity
- Na : 0.5 ppm, K : 0.5 ppm, Li : 0.5 ppm, Ca : 15 ppm, Ba : 50 ppm
- Filter Selection
- Automatic by soft touch key
- Resolution
- 0.1 ppm
- Detector
- Silicon Photodiode
- Linearity
- Better than 2%
- Unit of Measurement
- ppm, mg/L, meq/L and mmol/L
- Calculation of Concentration
- In built with system & can be measure by Sample Weight & Dilution Factor.
- Reproducibility
- Stability
- Curve Fitting Accuracy
- Correlation Factor
- Possible with System
- Quick Calibration Facility
- With Highest standard
- Calibration
- Up to 05 standards per Element with Curve Fitting Software in built with Instrument. (Automatic calculation of curve coefficient)
- Measurement of Sample
- Multi elements in single aspiration OR one by one.
- Calibration of Sample
- Multi elements in single aspiration OR one by one.
- Ignition System
- Auto Igniter
- Auto LPG Shut-off at Power Off
- Safety feature in case of sudden power failure
- Flame Type
- LPG & Dry Oil Free Air
- Auto Flame ON / OFF Detection
- Audio and Visual Alarm
- Gas Control
- Adjustable with Regulator
- Atomizer
- Axial Flow Type
- Air Flow Monitoring Facility
- 4 – 6 ml/min., In-built with Instrument
- Error Messages
- “LPG is Off”, “Home Detection”, “Memory Full”, “Flame is Off”, “Do Warm-up”, “In Correct Standard” & “Enter in Range”.
- Display
- 20 x 4 Line Alphanumeric Backlit LCD
- Keypad
- Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
- Air Supply Unit (Separate Unit)
- Consist of Air Compressor, Pressure Gauge, Pressure Regulator, Moisture Filter, PU Tube (2 Meters) and Air filter with moisture absorbent catalyst.
- Calibration Data Storage
- Non-Volatile Memory which can store, print & view last calibration data, calibration date & time, company name, calibration done by, calibration standard etc.
- Analysis Data Storage
- More than 1500 analysis data can be stored with corresponding calibration data, analysis date & time, record number, Batch number / ID, sample name, username, assigned method number, Corresponding Calibration data of lower & higher standard, Sample Result, Dilution factor, Sample Weight, Calculation of Concentration, instrument ID, serial number, model number, make, etc. Secondly, analysis data can be print online without saving in Instrument.
- Searching of data
- With record number, batch number, date, date range and by scrolling.
- Report Formation
- As per GLP Compliance
- 21 CFR Part-11 Compliance :
● In-built with system for ease of operation without PC.
● 2 Admin, 10 Reviewers, 40 Users & 1 Manufacturer with individual Login ID & Passcode.
● Audit trail can store more than 5000 lines in instrument.
● Audit trail includes action taken, date & time, action taken by whom.
● Audit trail can be transfer to PC through ANALABsafe software and can print on 40 / 80 column dot matrix printer for record.
● Admins can set up operation permission of each Users & Reviewers.
● Strict Admins control to audit and monitor usage and access.
● Audit trail record can be printed with Date, Date Range and by individual User’s ID.
● Once Audit trail transfer to the PC or Printer, system will delete Audit trail records and before final deletion, the system will give warning twice.
● Data integrity with database capability to store Analysis Data, Calibration Data, Calibration and Analysis Method, Audit trail and Passcode in instrument.
● No direct access to data from outside the instrument, thereby creating a closed system.
● Audit traceability to ensure that the device and results are appropriately handled.
● Auto User’s logout after defined Logout time.
● Auto Users & Reviewers block when password is entered incorrectly 3 consecutive times
- Printer Interface
- Centronics Parallel Port / Serial (RS-232 C) Port
- Printer
- 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
- PC Interface
- USB Type - B
- PC Connectivity
- By USB Interface Cable with Driver, ANALABsafe software to transfer Analysis data in GLP format, Calibration data and Audit trail from Instrument to PC in non-editable format (Cannot Delete)
- 230 VAC ± 10%, 50 Hz
- Dimensions:
● Net Dimension: 37 x 29 x 34 cm (L x W x H)
● Gross Dimension: 48 x 43 x 42 cm (L x W x H)
- Weight:
● Net Weight: 8.5 kg
● Gross Weight: 14 kg