- Element Range
- Na, K, & Li : 1-100 ppm, Ca : 15-100 ppm
- Sensitivity
- Na: 0.5 ppm, K : 0.5 ppm, Li : 0.5ppm, Ca : 15 ppm
- Filter Selection
- Automatic by soft touch key
- Detector
- Silicon Photodiode
- Linearity
- Better than 2%
- Unit of Measurement
- ppm
- Reproducibility
- Calibration
- Single Standard
- Ignition System
- Manual
- Flame Type
- LPG & Dry Oil Free Air
- Gas Control
- Adjustable with regulator
- Atomizer
- Axial Flow Type
- Error Messages
- “Home Detectionâ€
- Display
- 16 x 2 Line Alphanumeric LCD with Backlit
- Keypad
- Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
- Air Supply Unit (Separate Unit)
- Consist of Air Compressor, Pressure Gauge, Pressure Regulator, Moisture Filter, PU Tube (2 Meters) and Air filter with moisture absorbent catalyst.
- Calibration Data Storage
- Non-Volatile Memory, which can store, last calibration data for only one element.
- Power Supply
- 230 VAC ±10%, 50 Hz
- Dimensions:
â— Net Dimension: 37 x 29 x 34 cm (L x W x H)
â— Gross Dimension: 48 x 43 x 42 cm (L x W x H)
- Weight:
â— Net Weight: 8.5 kg
â— Gross Weight: 14Â kg