- pH Range
- -2.000 to 20.000 pH
- Selectable Resolution
- 0.1 / 0.01 / 0.001 pH
- Relative Accuracy
- ±0.02 / 0.002 pH
- mV/ORP Range – mV/RmV
- ±2000.0
- ORP
- EH auto calibration to standard hydrogen electrode
- Resolution
- 0.1
- Relative Accuracy
- ±0.2 mV or 0.05%, whichever is greater.
- Slope
- 80 - 120 %
- Auto Buffer Recognition :
1) 6.86 / 7.00 & 4.01 / 9.18 / 10.01 pH @ 25ºC for two points
2) 4.01, 6.86 / 7.00 & 9.18 / 10.01 pH @ 25ºC for three points
- Calibration Reminder Alarm
- 0-99 hours
- Result
- Auto / Manual
- pH Buffer Standard
- US / NIST / EURO / DIN
- Short Link
- Provided to diagnose the problem either with Instrument OR Electrode.
- Conductivity Range :
◠0-20.00, 20-200.0, 200-2000 µS/cm, 2-20.000, 20-200, 200-2000 mS/cm for K=1.0
◠0-10.00, 10-100.0, 100-1000 µS/cm, 1-10.000, 10-100, 100-1000 mS/cm for K=0.5
◠0-2.00, 2-20.0, 20-200, 200-2000 µS/cm, 2-20, 20-200 mS/cm for K=0.1
(Range value change depending upon cell selection)
- Conductivity Resolution
- 0.001, 0.01, 0.1 µS/cm & 0.001, 0.01, 0.1, 1 mS/cm
- TDS Range : 0 to 10, 100, 1000, 10000 ppm, 100ppt, 1000ppt
(Depending on TDS Factor)
- TDS Resolution
- 0.001, 0.01, 0.1 ppm & 0.001, 0.01, 0.1 ppt (1ppt=1000ppm)
- Number of Range
- Six
- Calibration Standard
- Six Conductivity & TDS Standard - One Per Range
- Calibration Factor
- Six – One Per Range
- Range Selection
- Auto / Manual
- Suitable Cell Constant
- 0.1, 0.5 & 1.0 K
- Adjustable Cell Constant
- 0.09 to 0.11 K, 0.45 to 0.55 K & 0.90 to 1.1 K
- Conductivity to TDS Conversion
- 0.40000 to 1.00000 (Adjustable Factor)
- Temperature Coefficient
- Six - 0-10.000 % per °C (Adjustable) – One Per Range
- Reading
- Conductivity at actual temperature and at 25°C & TDS with scrolling display.
- Temperature Range
- -5.0 to 130.0°C
- Temperature Resolution
- 0.1°C
- Temperature Compensation
- Auto / Manual
- Temperature Calibration
- In-built software and can be calibrated by users
- Calibration & Analysis
- pH Mode, ORP Mode, Conductivity Mode and TDS Mode.
- Error Message
- “Wrong standard / Cellâ€, “Cell Shortâ€, “Cell Openâ€, “Try Upper Rangeâ€, “Try Lower Rangeâ€, “Bad Buffer / Electrodeâ€, “pH Over Rangeâ€, “Temperature Probe Errorâ€, “In-complete Calibrationâ€, “Calibration Dueâ€, & “Low Memoryâ€.
- Display
- 16 x 2 Line Alphanumeric Backlit LCD.
- Keypad
- Membrane waterproof polyester soft keypad compliance to UL/ROHS.
- Calibration Data Storage
- Non-Volatile Memory can store, print & view calibration details.
- Analysis Data Storage : More than 500 analysis data which can be viewed & Print, company name, analysis date & time, record number, *Username, *Sample Name, *Bath Number/ID, pH, mV, temperature reading, calibration data with date & time, conductivity at 25°C and conductivity at actual temperature, % Tempco, conductivity to TDS conversion factor, TDS, instrument serial number, Instrument Make & Model & *Signature etc. Secondly, analysis data can be print online without saving analysis data in Instrument.
(*: Not possible to enter data by instrument keypad but it will be handwritten)
- Searching of data
- With record number, batch number, date, date range and by scrolling.
- Report Formation
- As per Non-GLP Compliance
- Printer Interface
- Centronics Parallel Port / Serial (RS-232 C) Port
- Printer
- 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
- Power Supply
- DC Adaptor
- Dimensions:
â— Net Dimension: 22 x 21 x 11 cm (L x W x H)
â— Gross Dimension: 32 x 29 x 23 cm (L x W x H)
- Weight:
â— Net Weight: 1.5 kg
â— Gross Weight: 3.5Â kg