- Wavelength Range
- 190-1100nm
- Spectral Bandwidth
- 1 nm
- Optical System
- Double Beam, Blazed Holographic Grating (1200 lines/mm)
- Wavelength Accuracy
- ±0.5nm
- Wavelength Repeatability
- ≤0.2nm
- Wavelength Setting
- Auto, Resolution 0.1nm
- Photometric Range
- 0~200%T,-4~4A,0~9999C
- Photometric Accuracy
- ±0.002 A (0~0.5A), ±0.003A (0.5~1A), ±0.3%T (0~100%T)
- Photometric Repeatability
- ≤0.001 A (0~0.5A), ≤0.002A (0.5~1A), ≤0.2%T (0~100%T)
- Stray Light
- ≤0.05%T(220/360nm)
- Scan Speed
- High, Medium, Low. Max.2000nm/minute
- Baseline Flatness
- ±0.001A/h (500nm,0A)
- Stability
- ±0.001A/h (500nm,0A)
- Noise
- ≤0 . 2%T / 3mi n ( 2 5 0 / 5 0 0 nm, 0%T ); ≤0 . 3%T / 3min (250/500nm,100%T)
- Sample Compartment
- 10mm Pathlength Cuvette
- Detector
- Silicon Photodiode
- Lamps
- Tungsten Lamp & Deuterium Lamp (Pre-aligned)
- Display
- Graphic LCD (320*240 Dots)
- Keypad
- 30-key Alphanumeric Membrane Keypad
- Output Port
- USB Port
- Printer
- Mini Serial Printer; PC Printer
- PC Software
- PC Scanning Software
- Power Requirements
- AC 90-250V, 50/60Hz
- Dimension
- 635x515x255mm
- Weight
- 26kg
- Holder
- 8 cell holder