- Conductivity Range :
◠0-20.00, 20-200.0, 200-2000 µS/cm, 2-20.000, 20-200, 200-2000 mS/cm for K=1.0
◠0-10.00, 10-100.0, 100-1000 µS/cm, 1-10.000, 10-100, 100-1000 mS/cm for K=0.5
◠0-2.00, 2-20.0, 20-200, 200-2000 µS/cm, 2-20, 20-200 mS/cm for K=0.1
(Range value change depending upon cell selection)
- Conductivity Resolution
- 0.001, 0.01, 0.1 µS/cm & 0.001, 0.01, 0.1, 1 mS/cm
- TDS Range : 0 to 10, 100, 1000, 10000 ppm, 100ppt, 1000ppt
(Depending on TDS Factor)
- TDS Resolution
- 0.001, 0.01, 0.1 ppm & 0.001, 0.01, 0.1 ppt (1ppt=1000ppm)
- Resistivity Range
- 20.00 MΩ to 50.00 KΩ, 5.00 KΩ, 500.00 Ω, 50.00 Ω, 0.50 Ω
- Resistivity Resolution
- 0.001
- Salinity
- 0.01 to 80000.0 ppm NaCl
- Temperature Range
- -5.0 to 130.0°C
- Temperature Resolution
- 0.1°C
- Temperature Compensation
- Auto / Manual
- Temperature Coefficient
- Six - 0-10.000 % per °C (Adjustable) – One Per Range
- Temperature Calibration
- In-built software and can be calibrated by users
- Number of Range
- Six
- Calibration Standard
- Six Conductivity & TDS Standard - One Per Range
- Calibration Factor
- Six – One Per Range
- Range Selection
- Auto / Manual
- Suitable Cell Constant
- 0.1, 0.5 & 1.0 K
- Adjustable Cell Constant
- 0.09 to 0.11 K, 0.45 to 0.55 K & 0.90 to 1.1 K
- Conductivity to TDS Conversion
- 0.40000 to 1.00000 (Adjustable Factor)
- Reading
- Conductivity at actual temperature and at 25°C, TDS, Resistivity & Salinity with scrolling display.
- TDS Calibration & Analysis
- In Separate Mode
- Error Messages
- “Wrong Standard / Cellâ€, “Cell Shortâ€, “Cell Openâ€, “Try Upper Rangeâ€, “Try Lower Rangeâ€, “Temperature Probe Error†& “Low Memoryâ€.
- Display
- 20 x 4 Line Alphanumeric Backlit LCD.
- Keypad
- Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS.
- Calibration Data Storage
- Non-Volatile Memory which can store, print & view with calibration date & time, cell constant, calibration range, standard conductivity solution, measured conductivity at 25°C, actual temperature, calibration factor, instrument serial number, instrument ID, model number etc.
- Analysis Data Storage
- More than 1500 analysis data can be stored with corresponding calibration data with company name, analysis date & time, record number, batch number / ID, sample name, username, conductivity at 25°C and conductivity at actual temperature, % Tempco, conductivity to TDS conversion factor, TDS, resistivity, salinity, instrument serial number, instrument ID, model number etc. Secondly, analysis data can be print online.
- Searching of data
- With record number, batch number/ID, date, date range and by scrolling.
- Report Formation
- As per GLP Compliance
- 21 CFR Part-11 Compliance :
â— In-built with system for ease of operation without PC.
â— 2 Admin, 10 Reviewers, 40 Users & 1 Manufacturer with individual Login ID & Passcode.
â— Audit trail can store more than 5000 lines in instrument.
â— Audit trail includes action taken, date & time, action taken by whom.
â— Audit trail can be transfer to PC through ANALABsafe software and can print on 40 / 80 column dot matrix printer for record.
â— Admins can set up operation permission of each Users & Reviewers.
â— Strict Admins control to audit and monitor usage and access.
◠Audit trail record can be printed with Date, Date Range and by individual User’s ID.
â— Once Audit trail transfer to the PC or Printer, system will delete Audit trail records and before final deletion, the system will give warning twice.
â— Data integrity with database capability to store Analysis Data, Calibration Data, Audit trail and Passcode in instrument.
â— No direct access to data from outside the instrument, thereby creating a closed system.
â— Audit traceability to ensure that the device and results are appropriately handled.
◠Auto User’s logout after defined Logout time.
â— Auto Users & Reviewers block when password is entered incorrectly 3 consecutive times.
- Printer Interface
- Centronics Parallel Port / Serial (RS-232 C) Port
- Printer
- 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
- PC Interface
- USB Type - B
- PC Connectivity
- By USB Interface Cable with Driver, ANALABsafe software to transfer Analysis data in GLP format, Calibration data and Audit trail from Instrument to PC in non-editable format (Cannot be Deleted)
- Power Supply
- DC Adaptor
- Dimensions:
â— Net Dimension: 22 x 21 x 11 cm (L x W x H)
â— Gross Dimension: 32 x 29 x 23 cm (L x W x H)
- Weight:
â— Net Weight: 1.5 kg
â— Gross Weight: 3.5Â kg